Chapter 13
Data analysis tools
9 Enter the 4-digit rotor number.
The rotor type for the entered rotor
number will be displayed.
Rotor type
Instruction
10 Verify that the single-stained sample tubes are placed at each rotor position
as described in the instruction area on the screen.
Caution
Place each sample tube in the correct position of the rotor so that it matches the instruction area on the
screen.
There is a risk of incorrect compensation results if the placement is incorrect.
11 Place the rotor in the auto loader.
For more information about how to place the rotor in the auto loader, see the following:
(➤P.166 "Chapter 12: 12.3.1 Running samples with the auto loader (auto loader analysis)", steps 2 to 3, and
steps 6 to 9)