XF 1600 IFU 5.0 en.pdf


Vista previa del archivo PDF xf-1600ifu50en.pdf


Página 1...194 195 196197198313

Vista previa de texto


Chapter 13

Data analysis tools

15 Repeat steps 13 and 14 for all the selected channels.
When you click [Next] in the last channel, the complete compensation matrix table is displayed.
Check the spillover values in the matrix table.

16 Click [OK] once you complete the compensation of all required channels.
The dialog box on the right appears.
If you want to redefine the gates, click [Cancel].

17 Click [OK].
The [Compensation Wizard] is completed.
The result is applied to the workitem or the subset panel activated in step 2.

Note:
If the auto-compensation is for a subset panel, make sure to click
The changes will be lost if you do not click
(Save).

(Save) to save the matrix table.

196/312
XF-1600 Instructions for Use