Data analysis tools
⏺ Performing automatic compensation using measured workitems of
single-stained samples
1 Confirm that workitems for the single-stained samples are included in the
worklist with the [Measured] status.
Prior to auto-compensation, you can register the workitems and run the single-stained samples for the desired
channels/fluorochromes.
For details on registering workitems, see the following:
(➤P.87 "Chapter 10 Registering panel/subset panel and workitems")
For details on running samples, see the following:
(➤P.161 "Chapter 12 Running samples")
2 Select and activate the workitem or subset panel to which you want to apply
spillover values from this auto-compensation.
3 Click the [Compensation] tab in the ribbon menu.
4 Click [Auto] in the [Compensation] group.
The compensation wizard starts and the
[Configuration Page] screen is displayed.
[Configuration Page] screen
197/312
XF-1600 Instructions for Use