XF 1600 IFU 5.0 en.pdf


Vista previa del archivo PDF xf-1600ifu50en.pdf


Página 1...301 302 303304305313

Vista previa de texto


Chapter 18 Troubleshooting

[Particle detection error]
Probable cause

Action

Concentration of the quality control material is too high or low,
or hardware-related errors exist.

Click [Accept] in the [Help] dialog box and check the quality
control material. Prepare the quality control material as
needed and rerun the material.
• If the error persists
Execute cleaning. For details, see the following:
(➤P.271 "Chapter 17: 17.6 Cleaning")
• If the error still persists
Perform baseline setup again then re-run daily QC.
For baseline setup, see the following:
(➤P.69 "Chapter 9: 9.8 Performing baseline setup")
For performing daily QC, see the following:
(➤P.73 "Chapter 9: 9.9 Performing daily QC")
• If the error still occurs after you have taken all the above
actions, contact your authorized local Sysmex
representative.

[QC material has expired]
Probable cause

Action

QC material has expired.

Click [Accept] in the [Help] dialog box.
Use a new QC bead lot to execute QC analysis.
When using a new lot, the lot information must be registered in
a new QC file.

Errors related to analysis
[Press Manual analysis start switch]
Probable cause

Action

The set time (2 hours) has elapsed since the XF-1600 entered
standby state.

Press the manual analysis start switch.

303/312
XF-1600 Instructions for Use