Chapter 18 Troubleshooting
[Particle detection error]
Probable cause
Action
Concentration of the quality control material is too high or low,
or hardware-related errors exist.
Click [Accept] in the [Help] dialog box and check the quality
control material. Prepare the quality control material as
needed and rerun the material.
• If the error persists
Execute cleaning. For details, see the following:
(➤P.271 "Chapter 17: 17.6 Cleaning")
• If the error still persists
Perform baseline setup again then re-run daily QC.
For baseline setup, see the following:
(➤P.69 "Chapter 9: 9.8 Performing baseline setup")
For performing daily QC, see the following:
(➤P.73 "Chapter 9: 9.9 Performing daily QC")
• If the error still occurs after you have taken all the above
actions, contact your authorized local Sysmex
representative.
[QC material has expired]
Probable cause
Action
QC material has expired.
Click [Accept] in the [Help] dialog box.
Use a new QC bead lot to execute QC analysis.
When using a new lot, the lot information must be registered in
a new QC file.
Errors related to analysis
[Press Manual analysis start switch]
Probable cause
Action
The set time (2 hours) has elapsed since the XF-1600 entered
standby state.
Press the manual analysis start switch.