Chapter 7
Instrument specifications
7.1.2 Performance
Item
Specification
Acquisition rate
50,000 events/second
Minimum detectable particle size
FSC: 0.5 μm, SSC: 0.5 μm
Fluorescence sensitivity
FITC < 100 MESF
PE < 50 MESF
APC < 100 MESF
Fluorescence linearity
FL1 R2 ≥0.97
FL2 R2 ≥0.97
FL6 R2 ≥0.97
Repeatability*
Intermediate
precision**
FSC and SSC sensitivity
CV (%) ≤10 %
Fluorescence sensitivity
CV (%) ≤10 %
WBC events
CV (%) ≤10 %
FSC and SSC sensitivity
CV (%) ≤10 %
Fluorescence sensitivity
CV (%) ≤10 %
WBC events
Carryover
CV (%) ≤10 %
<0.1 %
* Refers to the coefficient of variation when the measurement is repeated at least 10 times, using the following
samples and the measurement conditions (panels).
- For FSC/SSC and fluorescence sensitivity: CyFlow DailyQC by applying the panel [QC panel]
- For WBC events: Commercially available stabilized blood materials prepared at WBC 300-600/μL which are
qualified for use on the XF-1600, by applying the panel [WRP_R panel]
** Refers to the coefficient of variation when the measurement is repeated for 20 days by conducting 2 runs with
2 replicates each day in accordance with CLSI EP05-A3.
- For FSC/SSC and fluorescence sensitivity: CyFlow DailyQC by applying the panel [QC panel]
- For WBC events: Commercially available stabilized blood materials prepared at WBC 300-600/μL which are
qualified for use on the XF-1600, by applying the panel [WRP_R panel]
38/312
XF-1600 Instructions for Use