Chapter 18
Troubleshooting
[Aspiration unit left-right motor error]
Probable cause
Action
The aspiration unit left-right motor has malfunctioned.
Click [Execute] in the [Help] dialog box.
An aspiration unit motor test is executed.
If the error does not clear, the instrument requires servicing.
Contact your authorized local Sysmex representative.
[Tube holder motor error]
Probable cause
Action
The sample tube holder motor has malfunctioned.
Remove the sample tube, and then click [Execute] in the
[Help] dialog box.
A tube holder motor test is executed.
If the error does not clear, the instrument requires servicing.
Contact your authorized local Sysmex representative.
Errors related to the sample holder
[Tube remains in tube holder]
Probable cause
Action
• There was a tube remaining in the sample tube holder when
the instrument started up.
• There was a tube remaining in the sample tube holder when
the analysis mode was switched to manual analysis mode.
Remove the remaining tube from the tube holder and click
[Accept].
[Two tubes are in tube holder]
Probable cause
Action
Both a regular sample tube and a micro collection tube are set
for manual analysis.
Remove the sample tube that will not be run, and then click
[Accept] in the [Help] dialog box.
[No tubes are in tube holder]
Probable cause
Action
No sample tube is placed in the sample tube holder.
Click [Accept] in the [Help] dialog box.
Place a sample tube.
Errors related to the orders
[Unmeasured Workitem not active]
Probable cause
Action
No unmeasured workitem is activated at the time of manual
analysis.
Click [Accept] in the [Help] dialog box.
Activate the workitem you want to run and then execute
analysis.