Chapter 18 Troubleshooting
[Water leak 1 detected (analysis not possible)], [Water leak 2 detected (analysis not possible)]
Probable cause
Action
There is a liquid leak inside the analyzer.
Turn off the main power switch on the analyzer and wipe off
the liquid if the instrument is wet.
The instrument requires servicing. Contact your authorized
local Sysmex representative.
[Leak sensor 1 error], [Leak sensor 2 error]
Probable cause
Action
The liquid leak sensor has malfunctioned.
Turn off the main power switch on the analyzer.
The instrument requires servicing. Contact your authorized
local Sysmex representative.
[Rinse cup valve error], [Drain chamber 1 pinch valve error],
[Drain chamber 2 pinch valve error]
Probable cause
Action
The pinch valve has malfunctioned.
Click [Execute] in the [Help] dialog box.
A pinch valve test is executed.
If the error does not clear, the instrument requires servicing.
Contact your authorized local Sysmex representative.
Errors related to motors
[Sample aspiration motor error]
Probable cause
Action
The sample aspiration motor has malfunctioned.
Click [Execute] in the [Help] dialog box.
A sample aspiration motor test is executed.
If the error does not clear, the instrument requires servicing.
Contact your authorized local Sysmex representative.
[Sheath motor 1 error], [Sheath motor 2 error]
Probable cause
Action
The sheath motor has malfunctioned.
Click [Execute] in the [Help] dialog box.
A sheath motor test is executed.
If the error does not clear, the instrument requires servicing.
Contact your authorized local Sysmex representative.
[Aspiration unit up-down motor error]
Probable cause
Action
The aspiration unit up-down motor has malfunctioned.
Turn off the main power switch on the analyzer.
The instrument requires servicing.
Contact your authorized local Sysmex representative.
299/312
XF-1600 Instructions for Use