Chapter 18 Troubleshooting
18.3
Causes of errors and remedial actions
When an error occurs, the [Help] dialog box appears. Refer to the causes and actions described below and take the
appropriate action.
If there is more than one possible cause, take action in the order.
Risk of infection
When taking remedial action to handle an error, always wear adequate personal protective equipment such
as protective gloves, a protective mask, protective eyewear, and a lab coat. Wash your hands with
antiseptic solution after completing the task.
There is a risk of infection.
Errors related to pressure
[Abnormal pressure loss]
Probable cause
Action
The air pump has malfunctioned.
Click [Execute] in the [Help] dialog box.
An air pump test is executed.
[0.06MPa pressure error]
Probable cause
Action
The 0.06 MPa relief valve has malfunctioned.
Click [Execute] in the [Help] dialog box.
An air pump test is executed.
[-0.03MPa pressure error]
Probable cause
Action
The -0.03 MPa relief valve has malfunctioned.
Click [Execute] in the [Help] dialog box.
An air pump test is executed.
Errors related to temperature
[Ambient temperature error (high temperature)], [Ambient temperature error (low temperature)]
Probable cause
Action
The ambient temperature (environmental temperature) is
either:
• >38 °C (high temperature error)
• < 5 °C (low temperature error)
Click [Execute] in the [Help] dialog box.
Check [Environment Temp.] in the [Sensor Display] (Sensor 1)
dialog box that appears and adjust the room temperature.
(➤P.281 "Chapter 17: 17.12 Checking the instrument status
(sensors)")
297/312
XF-1600 Instructions for Use