Chapter 17
Maintenance
17.9.2 Pinch valve test
Pinch valves are located in the waste chamber and rinse cup aspiration unit.
1 Display the maintenance menu.
(➤P.264 "17.2 Displaying the maintenance menu")
2 Click [Pinch Valve Test].
The maintenance name appears in the status area. If you click it, the dialog box appears. Wait until the process
finishes. When the test finishes normally, the dialog box closes automatically.
17.9.3 Sample aspiration motor test
This tests sample aspiration and dispensing by the sample aspiration motor.
1 Display the maintenance menu.
(➤P.264 "17.2 Displaying the maintenance menu")
2 Click [Sample Aspiration Motor Test].
The maintenance name appears in the status area. If you click it, the dialog box appears. Wait until the process
finishes. When the test finishes normally, the dialog box closes automatically.
17.9.4 Sheath motor test
This tests delivery of the sample to the FCM detector by the sheath motor.
1 Display the maintenance menu.
(➤P.264 "17.2 Displaying the maintenance menu")
2 Click [Sheath Motor Test].
The maintenance name appears in the status area. If you click it, the dialog box appears. Wait until the process
finishes. When the test finishes normally, the dialog box closes automatically.
17.9.5 Aspiration unit motor test
This tests moving of the sample aspiration pipette by the aspiration unit motor.
1 Display the maintenance menu.
(➤P.264 "17.2 Displaying the maintenance menu")