Chapter 17

17.10

Maintenance

Replenishing the reagent

If an incorrect reagent is accidentally connected, the reagent in the fluid line should be replenished.

1 Make sure that the correct reagent is connected.
2 Display the maintenance menu.
(➤P.264 "17.2 Displaying the maintenance menu")

3 Click [Reagent Replenishment].
The dialog box on the right appears.

[Reagent Replenishment] dialog box

4 Select the checkbox of the reagent you want to replenish and click [Execute].
Reagent replenishment starts. Progress is shown in the progress bar on the screen. Wait until the process
finishes. When the process finishes, the dialog box closes automatically.

278/312
XF-1600 Instructions for Use