Chapter 17
Maintenance
2 Click [Aspiration Unit Motor Test].
The maintenance name appears in the status area. If you click it, the dialog box appears. Wait until the process
finishes. When the test finishes normally, the dialog box closes automatically.
17.9.6
Tube holder motor test
This tests the moving of a sample tube holder by the tube holder motor.
1 Make sure that the analysis mode of the status area is [Manual]. If the mode is
set to [AutoLoader], press the mode switch.
The mode changes to [Manual] and the sample tube holder is automatically ejected.
2 Display the maintenance menu.
(➤P.264 "17.2 Displaying the maintenance menu")
3 Click [Tube Holder Motor Test].
The maintenance name appears in the status area. If you click it, the dialog box appears. Wait until the process
finishes. When the test finishes, the dialog box closes automatically.
17.9.7
Auto loader motor test
This tests the moving of the auto loader by the auto loader motor.
1 Make sure that the analysis mode of the status area is [AutoLoader]. If the
mode is set to [Manual], press the mode switch.
The mode changes to [AutoLoader] mode.
2 Make sure that the auto loader is closed.
3 Display the maintenance menu.
(➤P.264 "17.2 Displaying the maintenance menu")
4 Click [Auto Loader Motor Test].
The maintenance name appears in the status area. If you click it, the dialog box appears. Wait until the process
finishes. When the test finishes normally, the dialog box closes automatically.