Chapter 17

Maintenance

4 Click [Rinse Flowcell].
The dialog box on the right appears.

[Rinse Flowcell] dialog box

5 Dispense CyFlow CLEAN into a sample tube.
The amount of CyFlow CLEAN required for rinsing the flowcell is approximately 2.5 mL. Make sure that CyFlow
CLEAN has not expired.

Warning
CyFlow CLEAN is a strong alkali. Do not mix it with any acidic substances.

Caution
• When handling CyFlow CLEAN, always wear adequate personal protective equipment such as protective
gloves, a protective mask, protective eyewear, and a lab coat.
• Take care that the fluid does not spill.

6 Place CyFlow CLEAN in the sample tube holder.
Place it in the front holder, when you face the
analyzer.
Make sure that the tube is uncapped.

274/312
XF-1600 Instructions for Use