Chapter 12

Running samples

[Laser]

Indicates the status of each laser. When lasers are powered on, each
button displays the laser color as follows:
• 488 nm (blue)*
• 638 nm (red)
• 405 nm (violet)
Before running a sample, set the ON/OFF status of each laser by clicking
the colored buttons. When the laser is powered off, the color changes to
gray and the corresponding channels are also disabled.
* The 488 nm (blue) laser cannot be turned OFF because it covers FSC/
SSC channels that are required for data acquisition, regardless of FL
channels selected.

12.2.2 Setting stop conditions
Stop conditions allow you to specify when the collection of data will end.
This procedure cannot be performed while measurement is in progress. Display and set the desired workitem or
subset panel before measurement.

Note:
This operation is not available if the user’s role is [Operator].

1 Select the workitem or subset panel you want to run, and activate it by
clicking

(Active).

Note:
The measurement condition setting area and the workarea will change for the active workitem or subset
panel.

2 Click the [Stops] tab in the measurement condition setting area.
The screen on the right appears.
The screen shown on the right is
displayed when [Single display] has been
selected for the [Display] setting in the
IPU settings.
Screen displays and operation vary when
[Multi-display] is selected.
For further information on the [Display]
setting, see the following:
(➤P.238 "Chapter 15: 15.4.4 Display")
[Stops] tab (when set to [Single display])

164/312
XF-1600 Instructions for Use