Chapter 12
Running samples
12.2.1 Setting acquisition conditions
Set the particle acquisition conditions in the measurement condition setting area.
• Channel settings
- Channels to be acquired, the name of each channel, and the parameters (Height, Area, or Width)
- Threshold
- Trigger setting
- Gain/PMT voltage setting
• Flow rate
• Sample volume
• Laser ON/OFF (laser control)
Note:
• Only Threshold and Gain/PMT voltage values can also be changed during acquisition. The other
conditions cannot be changed once the acquisition begins.
• This operation is not available if the user’s role is [Operator].
1 Select the workitem or subset panel of the sample you want to run, and
activate it by clicking
(Active).
The measurement condition setting area and the workarea will change for the active workitem or subset panel.
2 Click the [Settings] tab in the measurement condition setting area.
The screen on the right appears.
The screen shown on the right is displayed when
[Single display] has been selected for the [Display]
setting in the IPU settings.
Screen displays and operation vary when [Multidisplay] is selected.
For further information on the [Display] settings, see
the following:
(➤P.238 "Chapter 15: 15.4.4 Display")
[Settings] tab (when set to [Single display])
3 Select the desired condition for each item below.
[LOGIC]
This defines the triggered logical condition. When more than one
parameter is listed, select either [OR] or [AND] to define the logical
condition to be triggered.
[OR]
Signals must be equal to or greater than any one of the set threshold
values to be acquired.
[AND]
Signals must be equal to or greater than all of the set threshold values to
be acquired.