Chapter 11

Panel QC

[Lower Limit]

Displays the lower limit for the selected item.

4 Enter the 4-digit rotor number in the [Rotor No] for the active subset panel.
The rotor type for the entered rotor number is displayed, and the [Start] button becomes active.

5 Prepare the required sample tubes for the active subset panel, and place them
in the rotor as described in [Rotor Pos. Information].
For each sample tube to be run for auto-compensation, prepare 200 µL of single-stained sample regardless of
the measurement conditions of the subset panel.

Caution
Place each sample tube in the correct position of the rotor so that it matches the [Rotor Pos. Information].
There is a risk of incorrect QC data if the placement is incorrect.

6 Place the rotor in the auto loader.
For more information on how to place the rotor in the auto loader, see the following:
(➤P.166 "Chapter 12: 12.3.1 Running samples with the auto loader (auto loader analysis)", steps 2 and 3, and
steps 6 to 9.)

7 Click [Start] at the upper right of the [Daily QC] dialog box.
Auto loader measurement starts.
Once the measurement of the active subset panel is complete:

⏺ When [Not apply] or [Apply Daily QC results] is selected in [QC result to
Measurement Setting]
• If the panel has only one subset panel, or the measurement was for the last subset panel, the [OK] button at
the lower right becomes active.
Click [OK] and go to step 10.
• If the panel has any unmeasured subset panels, the [Next] button appears at the upper right of the dialog box.
Click [Next], then the
in the [Active] column moves to the next subset panel.
Repeat steps 4 to 7 for the remaining subset panels.

⏺ When [Depends on Daily QC results] is selected in [QC result to
Measurement Setting]
Go to step 8.

146/312
XF-1600 Instructions for Use