Chapter 11
Panel QC
⏺ If you select [Not apply]
Click [OK] and the dialog box closes.
• If the panel has only one subset panel, or the measurement was for the last subset panel, the [OK] button at
the lower right of the [Daily QC] dialog box becomes active.
Click [OK] and go to step 10.
• If the panel has any unmeasured subset panels, the [Next] button appears at the upper right of the dialog box.
Click [Next], then the
in the [Active] column moves to the next subset panel.
Repeat steps 4 to 7 for the remaining subset panels.
⏺ If you select [Apply Daily QC results]
The [Select] button is activated. Go to step 9.
9 If needed, select additional subset panels to apply QC results.
1 Click [Select].
The following dialog box appears.
The panel/subset panels that were measured under the same measurement conditions (i.e., the selected
channels and Height/Area/Width) as the current subset panel are displayed in the [Panel List] and [Subset
Panel List] areas.
2 If needed, select additional subset panels to apply QC results.
• Select the desired panel and check [Gain/PMT] and/or [Comp.] of the subset panel as needed.
• Unchecked items are not applied.
148/312
XF-1600 Instructions for Use