Chapter 11

3 Select panel QC file and click

Panel QC

(Daily QC).

The [Daily QC] dialog box appears.

[Rotor No]

Enter the 4-digit number on the rotor barcode label.

[Rotor Type]

Displays the rotor type for the entered rotor number.

[Start]

Click to start the QC run by auto loader.

[QC File Name]

Displays the QC file name.

[Panel Name]

Displays the panel name registered in the QC file.

[Lot No.]

Displays the lot number for the QC file.

[Subset Panel]

Displays information about the subset panels registered in the QC file.
If the panel has multiple subset panels, the first subset panel is activated when
starting the daily panel QC. Once the run for the active subset panel is completed,
automatically moves to the next subset panel.
The screen displays information about the active subset panel.

[Expiration Date]

Displays the expiration date of the QC material.

[Comment]

Displays comments about the QC material.

[Rotor Pos. Information] Displays information about the tube loaded at each position in the rotor.
[Pos.]

Displays the position number of the tube in the rotor.

[Tube]

Displays information about the tube.

[Gain/PMT]

Baseline settings are displayed in the [Item] and [Target] columns based on the
selected limit setting. Daily panel QC results are displayed in the [Value] column.

[Compensation]

Baseline settings are displayed in the [Item] and [Target] columns based on the
selected limit setting. Daily panel QC results are displayed in the [Value] column.

[Statistics]

Baseline settings are displayed in the [Item] and [Target] columns based on the
selected limit setting. Daily panel QC results are displayed in the [Value] column.

[Upper Limit]

Displays the upper limit for the selected item.

145/312