Chapter 11
[Limit Input]
Histogram/Plot area
Panel QC
If you selected [User Setting] for [Limit], enter the lower and upper limit ranges in
the field to the right of the [-] and [+] marks respectively. By default, a numerical
value that is 10 % of the target value is displayed. Check the values and edit them
if necessary.
If you selected [User Value] for [Limit], enter the lower/upper limit and target values
in the fields next to the [~] mark. Enter the lower limit value in the leftmost entry
field and the upper limit value in the rightmost entry field. If you enter a lower/upper
limit value higher/lower than the target value, the field will be marked in red, and
you cannot save the setting.
Displays the plot or histogram of FSC/SSC for the gate in the selected row. If
multiple rows are selected, no plot or histogram is displayed.
2 If you want to select management items individually, click [Edit].
The [Select Statistics Item] dialog box is displayed.
Statistics list
Histogram/Plot area
[Gate Name]
Displays a tree-structure list of subset panel gates.
Statistics list
Displays the statistics items and the values that are calculated in the subset panel.
Histogram/Plot area
Displays a plot for the gate in the selected row.
3 If there is any item you want to remove from management, clear the checkboxes and click [OK].
4 Click the row you want to configure in the statistics list.
The background color of the selected row changes to blue, and a plot is displayed in the histogram/plot
area.
5 Select a limit value from the [Limit] drop-down menu.
If you select [User Setting], enter the [Upper Limit] and [Lower Limit].
6 Repeat sub-steps 4 and 5 to configure all the rows in the statistics list.
143/312
XF-1600 Instructions for Use