XF 1600 IFU 5.0 en.pdf


Vista previa del archivo PDF xf-1600ifu50en.pdf


Página 1...140 141 142143144313

Vista previa de texto


Chapter 11

Panel QC

2 Click the cell you want to configure in the compensation matrix.
The background color of the selected cell changes to blue, and a plot is displayed in the plot area.

3 Select a limit value from the [Limit] drop-down menu.
If you selected [User Setting], enter the [Upper Limit] and [Lower Limit].

4 Repeat sub-steps 2 and 3 to configure all the cells in the compensation
matrix.

⏺ Setting a statistics value as a QC item

1 Click the [Statistics] tab.
The screen below is displayed.
Statistics list

Limit setting area

Histogram/Plot area

Statistics list

Displays statistics items and the values for all gates in the subset panel.

[Gate]

Displays the names of all gates in the subset panel.

[Statistics]

Displays the statistics items for all gates in the subset panel.

[Lower Limit]/
[Upper Limit]

If you selected [User Setting] or [User Value] for [Limit], the configured limit values
are displayed.

[Target]

Displays the statistics values for all gates in the subset panel.

[Edit]

Click to display a dialog box that allows you to select or deselect a statistics item.

Limit setting area
[Limit]

Define the limit range for each statistics value.
Select the upper and lower limit settings from the drop-down menu. You can select
[±2SD], [±3SD], [User Setting], or [User Value]. The default setting is [±2SD].

142/312
XF-1600 Instructions for Use