XF 1600 IFU 5.0 en.pdf


Vista previa del archivo PDF xf-1600ifu50en.pdf


Página 1...138 139 140141142313

Vista previa de texto


Chapter 11

Panel QC

[Target]

Displays the Gain/PMT values from when the subset panel master was last saved.
This value cannot be changed.

[Add]

Click to add a channel to the lowermost row. If there are no channels that can be
added, the button cannot be clicked.

[Delete]

Click to delete the selected row.

Parent gate and limit
area

Select a desired parent gate and the limit for each channel.

[Parent Gate]

Select the parent gate from the drop-down menu.

[Limit]

Select the upper and lower limit setting from the drop-down menu. You can select
[±2SD], [±3SD], [User Setting], or [User Value]. The default setting is [±2SD].

[Limit Input]

If you selected [User Setting] for [Limit], enter the lower and upper limit ranges in
the field to the right of the [-] and [+] marks respectively. By default, a numerical
value that is 10 % of the target value is displayed. Check the values and edit them
if necessary.
If you selected [User Value] for [Limit], enter the lower/upper limit and target values
in the fields next to the [~] mark. Enter the lower limit value in the leftmost entry
field and the upper limit value in the rightmost entry field. If you enter a lower/upper
limit value higher/lower than the target value, the field will be marked in red, and
you cannot save the setting.

Histogram/Plot area

Displays particle data (number of events) of the selected channel as a histogram.
No histogram is displayed if multiple channels are selected.

[Histogram / Plot]

Define the peak range for the population you want to manage by dragging the
marker gate.

2 Add or delete channels as necessary.
• To add a row, click [Add]. If no row can be added, [Add] is disabled.
• To change the channel and/or parameter of a newly added row, click the [Channel] or [H/A/W] column
and change it as needed.
• To delete a row, select it and click [Delete].

3 Click the row you want to change and configure the following settings:
• [Parent Gate]:
• [Limit]:

Select the parent gate from the drop-down menu.
Select the limit value from the drop-down menu. If you select [User Setting], enter the
[Upper Limit] and [Lower Limit].
• [Histogram / Plot]: Drag the marker gate to set the peak range for the population you want to manage.
Repeat this step to configure all channels you want to manage.

140/312
XF-1600 Instructions for Use