XF 1600 IFU 5.0 en.pdf

Vista previa de texto
Chapter 11
Panel QC
4 Click [OK].
The dialog box closes.
The selected subset panel information is reflected in the [QC result to Measurement Setting] area of the
[QC File Registration / Setup] screen.
10 Set the baseline.
You can choose required QC items from the following three tabs:
• Gain/PMT
(➤P.139 "⏺ Setting the Gain/PMT value as a QC item")
• Compensation (➤P.141 "⏺ Setting the compensation values as a QC item")
• Statistics
(➤P.142 "⏺ Setting a statistics value as a QC item")
⏺ Setting the Gain/PMT value as a QC item
1 Click the [Gain/PMT] tab.
The screen below is displayed.
Channel list
Parent gate and limit area
Histogram/Plot area
Channel list
The channels and Gain/PMT values of the subset panel are displayed
automatically.
[Channel]
By default, the channels used for the plots of the subset panel are displayed.
Channels to be monitored by the panel QC process can be added or deleted.
[H/A/W]
You can select a parameter to be managed for each channel.
[H], [A], and [W] indicate Height, Area, and Width, respectively.
[Lower Limit]/
[Upper Limit]
If you selected [User Setting] or [User Value] for [Limit], the configured limit values
are displayed.
139/312
XF-1600 Instructions for Use
