Chapter 10
Registering panel/subset panel and workitems
11 Click
(Save).
Note:
• To save changed settings, click the
(Save) button. If you try to display a different screen without
saving the changes, a confirmation dialog box appears.
• You can back up measurement condition settings and the compensation matrix to an external file, and
apply them to a different subset panel or an unmeasured workitem.
For details, see the following:
(➤P.178 "13.4 [XF-1600 File] tab ⏺[Backup]")
10.4
Registering a workitem in the worklist
A workitem consists of associated panel/subset panel information, together with the status/results of
measurement and sample information, allowing you to manage the information collectively.
The workitem registration procedure depends on whether you are using the auto loader or the sample tube
holder to perform analysis.
For details on using the auto loader, see the following:
(➤P.98 "10.4.1 Registering a workitem in the worklist (for auto loader analysis)")
For details on using the sample tube holder, see the following:
(➤P.104 "10.4.2 Registering a workitem in the worklist (for manual analysis)")
Note:
• The worklist can contain up to 10,000 workitems if there is more than 50 GB of available space on the
hard disk drive.
• It is recommended to back up and delete old workitems periodically to avoid slowing down the IPU.
For details, see the following:
(➤P.178 "13.4 [XF-1600 File] tab ⏺[Backup]")
• If you try to register a new workitem exceeding the maximum number, a confirmation dialog box appears,
depending on the setting. If you click [OK] to proceed, the oldest workitem will be automatically deleted.
For details, see the following:
(➤P.244 "15.4.9 Worklist")
10.4.1 Registering a workitem in the worklist (for auto loader analysis)
1 Click the [Worklist] tab in the worklist area.
The worklist appears.