Chapter 9

Cytometer QC and setup

⏺ If the daily QC fails
An error message appears in the status area, and an alarm sounds depending on the setting.
If the results are displayed on the [Daily QC] dialog box, click [OK] to save the results after the progress bar is
closed.
Check the error by clicking the
.
For details, see the following:
(➤P.296 "Chapter 18: 18.2 [Help] dialog box")
If no results are displayed on the [Daily QC] dialog box, click [Cancel] after the progress bar is closed. Check
the error by clicking the
.
For details, see the following:
(➤P.296 "Chapter 18: 18.2 [Help] dialog box")

Note:
All alarms sounding on the IPU will stop when you press any key on the keyboard.
If any items fail, the daily QC run cannot be completed successfully. Before proceeding to run samples, re-run
the daily QC. If the error persists, execute cleaning and re-run the daily QC.
If the QC continues to fail, perform baseline setup again or contact your authorized local Sysmex
representative.