Chapter 8

8.3

Basic operation

Shutdown

Shutdown consists of rinsing the hydraulic system with CyFlow CLEAN and turning the instrument power off.
When you finish analysis work for the day, always perform a shutdown. If sample acquisition and analysis work will
continue for more than one day, perform a shutdown once per day.

1 Make sure that the analysis mode of the status area is set to [Manual]. If the
mode is set to [AutoLoader], press the mode switch.
The mode changes to [Manual] and the sample tube holder is automatically ejected.

2 Make sure that the instrument is in the ready state.
If the analysis status LED/instrument status shown in the
bottom left corner of the IPU screen is not green, wait until it
turns green.

Analysis status
indicator LED

3 Click [Maintenance] in the screen transition area.
The submenu on the right appears.