Data analysis tools
[Min]
Enter the minimum number of counts of the X-axis.
[Max]
Enter the maximum number of counts of the X-axis.
[Negative %]
When [Logicle] is selected as the scale type, the range can be set from 0.0
to 30.0 %. The number can be selected from the drop-down menu in
increments of 5 %.
When [Lin] or [Log] is selected as the scale type, this button is disabled.
[Y] Parameter
Name
Select a parameter of the Y-axis from the drop-down menu.
Type
Select the scale of the Y-axis from the drop-down menu. The available
scale types are [Lin] (Linear), [Log], and [Logicle].
[Min]
Enter the minimum number of counts of the Y-axis.
[Max]
Enter the maximum number of counts of the Y-axis.
[Negative %]
When [Logicle] is selected as the scale type, the range can be set from 0.0
to 30.0 %. The number can be selected from the drop-down menu in
increments of 5 %.
When [Lin] or [Log] is selected as the scale type, this button is disabled.
13.6.2 [Histogram Style] group
The [Histogram Style] group contains options for scaling the selected histogram for optimal plot visualization.
[Histogram Style] group
The [Histogram Style] group contains the following options.
[Auto Scaling]
Select this option to enable automatic scaling of the selected histogram.
[Max Counts]
Select this option to manually adjust the scales of the selected histogram.
Enter the desired number of maximum counts of the Y-axis in the [Max
Counts] field.
190/312