Chapter 12

Running samples

8 Verify that the pedestal is oriented properly.
An error message will be displayed at the start of
aspiration if the pedestal is not positioned in the
proper orientation.

Top view

9 Install the rotor as shown in the figure.
⏺ If using a fixed rotor
Orient the white line to match the
position shown in the figure on the
right.
The rotor will click into place when
properly positioned.

White line

Top view

Groove

If using a fixed rotor

⏺ If using a swing-out rotor
Position the rotor so that the
protrusions on the back fit into the
grooves on the pedestal. The rotor will
click into place when properly
positioned.

Position 1
Barcode

Make sure that the placement of
position 1 and the barcode label
matches the figure on the right.
Top view

Groove

If using a swing-out rotor

169/312
XF-1600 Instructions for Use