Chapter 12

Running samples

4 Prepare the rotor.
If connected to the PS-10, retrieve a rotor for which pre-processing has completed from the PS-10.
If not connected to the PS-10, prepare a rotor into which the tubes to be measured have been loaded. When
using a tube that does not fit the diameter of the rotor adapter, remove the adapter.

5 (If not connected to the PS-10) Make sure that the workitems for the tubes in
the rotor have been registered.
For details on registering a workitem in the worklist, see the following:
(➤P.98 "Chapter 10: 10.4.1 Registering a workitem in the worklist (for auto loader analysis)")

6 (If not connected to the PS-10) Make sure that the position of each sample
tube on the rotor matches the rotor position registered in the workitem.

Caution
Place each sample tube in the correct position of the rotor so that it matches the [Rotor Position] displayed
on the worklist.
There is a risk of incorrect data if the placement is incorrect.

7 Open the auto loader.

168/312