Chapter 11
Panel QC
⏺ Chart display area (left)
[Item Data]
Displays the names of panel QC parameters and the values from the daily panel QC
run selected by the cursor. When it is outside the limits, the item is highlighted in red.
[UL]
Displays the upper limit value based on the selected limit setting.
[Target]
If [User Setting] is applied for the range of limit, the target value from the baseline
setup is displayed.
If [±2SD] or [±3SD] is selected, the target value is calculated as follows:
- When the number of data point is 1, the target will be the value of the data point.
(The result will be [Pass].)
- When the number of data points is 2, the target will be the value of the first data
point. (If the value of the second data point is within the range of the first value ± 1
minimum significant digit, the result will be [Pass].)
- When the number of data points is 3 or more, the target will be the mean value of
the previous data points after the last baseline setup or QC file setting.
[LL]
Displays the lower limit value based on the selected limit setting.
⏺ Chart display area (right)
The results are plotted by time and displayed in the chart area as a broken-line graph.
Each data point on the chart corresponds to a single daily panel QC run and the type of points indicates
whether the result is within the limit (Pass) or outside the limits (Fail). It also indicates whether the data points
of a particular run are used for panel QC management.
Comment display area
Baseline
Comment marks
Vial line
Data points
Cursor
Date
Data points
Indicates that the QC value is within the upper and lower limits.
Indicates that the QC value is outside the upper or lower limit.
153/312
XF-1600 Instructions for Use