Chapter 11

11.8

Panel QC

Checking daily panel QC results

In the QC charts area, you can check the pass/fail status of daily panel QC results run against the baselines of a
selected panel QC file.

11.8.1

Radar charts

The most recent daily panel QC results of the selected subset panel of the panel QC file are displayed in radar
charts.

For example, the chart in the upper left corner
displays the data points of the following four
parameters.
• SSC-H_PMT
• FSC-H_Gain
• FL1-A_PMT
• FL2-A_PMT

Upper limit

Lower limit

Target

Data points

Radar chart
Item

Displays the name of each parameter. Any parameter exceeding the limit will be
highlighted in red.

Lower limit

Indicates the lower limit of the parameter.

Upper limit

Indicates the upper limit of the parameter.

Target

Indicates the target of the parameter.

Data points

Indicates the data points from the selected subset panel of the panel QC file.

151/312