Chapter 17

Maintenance

[Type]

Displays the type of barcode.

Reading result

Displays the reading result.
One of the following symbols is displayed, depending on the result:
If there was no problem with the reading, nothing is displayed.
[E]: Reading error or invalid check digit.
[+]: A value that is longer than the specified number of digits was read.
[-]: A value that is shorter than the specified number of digits was read.

10 Click [Cancel].
The dialog box closes.

11 Open the auto loader.
12 Remove the rotor.
13 Close the auto loader.

17.14

Replacing the air pump

The air pump must be replaced once every 500 hours of operating time, or once every two years, whichever occurs
first. Continuing to use an old air pump may result in air pump failure and other problems.
When [Replace air pump.] appears, promptly replace the air pump.
A replacement air pump is not included with the instrument. Please contact your authorized local Sysmex
representative to purchase a replacement air pump.

Caution, Hot
The surface of the air pump is hot. After turning off the main power switch on the analyzer, make sure that
the air pump has cooled sufficiently before replacing it.
Use the following air pump:
Product code
05104711

Item name
Air pump set No. 1

1 Shut down the instrument.
For the shutdown procedure, see Chapter 8.
(➤P.51 "Chapter 8: 8.3 Shutdown")

284/312
XF-1600 Instructions for Use