Chapter 17
17.12
Maintenance
Checking the instrument status (sensors)
You can check the pressure, temperature, laser currents, and sensor status of each unit. The displayed information is
updated every 0.5 seconds.
To display this information, you can also execute analysis with the [Sensor Display] dialog box open.
1 Display the maintenance menu.
(➤P.264 "17.2 Displaying the maintenance menu")
2 Click [Sensor Display].
The dialog box on the right appears.
Shortcut buttons
Next button
[Sensor Display] (Sensor 1) dialog box
Shortcut buttons
These buttons are used to display the dialog boxes that appear in the
[Sensor Display] dialog box.
Click a button of a dialog box that does not currently appear to display the
dialog box.
[Pressure (units: MPa)]
Displays the pressure of each unit.
[Temperature (units: °C)]
Displays the temperature of each instrument unit and the ambient
temperature.
[Laser Current, Laser Power]
Displays the output current and power of the laser.
Next button
Click to display the [Sensor Display] (Sensor 2) dialog box.
281/312