Chapter 17
Maintenance
Time guidelines for reagent replacement are shown below.
Reagent name
CyFlow SHEATH
17.6
Time
About 1 minute
Cleaning
Clean the fluid paths in the instrument using CyFlow CLEAN. Perform cleaning every 100 measurements or every
7 days.
An error message appears if the cleaning has not been performed beyond this period.
1 Make sure that the analysis mode of the status area is [Manual]. If the mode is
set to [AutoLoader], press the mode switch.
The mode changes to [Manual] and the sample tube holder is automatically ejected.
2 Display the maintenance menu.
(➤P.264 "Chapter 17: 17.2 Displaying the maintenance menu")
This step is not necessary if the [Cleaning] dialog box is already displayed.
Go to step 4.
3 Click [Cleaning].
The dialog box on the right appears.
[Cleaning] dialog box
4 Dispense CyFlow CLEAN into a sample tube.
The amount of CyFlow CLEAN required for executing cleaning is approximately 2.5 mL. Make sure that CyFlow
CLEAN has not expired.
Warning
CyFlow CLEAN is a strong alkali. Do not mix it with any acidic substances.
271/312
XF-1600 Instructions for Use