Chapter 17

Maintenance

Note:
If you are manually entering the reagent code, click the name
of the reagent to be replaced in the [Reagent Replacement]
dialog box. The dialog box on the right appears.
Select the [Replace the reagent] checkbox, enter the
[Reagent Code:], and click [OK]. Then follow the instructions
in the dialog box to register the reagent information.

4 Remove the cap from the new reagent container.
5 Remove the cap of the old reagent container.

6 Pull the spout set straight up and out.

Caution
Do not touch the aspiration nozzle on the spout set. Take care that dust does not get on the spout set.

7 Insert the spout set straight into the new reagent container and close the cap.
8 Click [Execute].
Reagent replacement starts. Wait until the process finishes. When the process finishes, the dialog box closes
automatically.

270/312
XF-1600 Instructions for Use