Chapter 13

Data analysis tools

11 Repeat steps 9 and 10 for all the selected channels.
When you click [Next] in the last channel, the complete [Compensation Matrix] is displayed.
Check the spillover values in the matrix.

12 Click [OK] after you complete the compensation of all required channels.
The dialog box on the right appears.
If you want to redefine the gates, click [Cancel].

13 Click [OK].
The [Compensation Wizard] is completed.
The result is applied to the workitem activated in step 2.

Note:
If the auto-compensation is for a subset panel, make sure to click
The changes will be lost if you do not click
(Save).

(Save) to save the matrix table.

13.7.2 [Apply] group
The [Apply] group contains the option to enable or disable compensation.

[Apply] group

[Enable Compensation]

Select the checkbox to enable compensation of the active workitem or
subset panel. When cleared, compensation is disabled. If you clear the
checkbox after you apply the compensation, the plots change to the status
before the compensation was applied. The spillover values you set in the
compensation matrix remain even after clearing the [Enable
Compensation] checkbox.

200/312
XF-1600 Instructions for Use