Chapter 11
Panel QC
⏺ Adding comments to the results
1 Move the cursor to the results of the date/time you want to add comments to.
2 Click
(Comment) in the command buttons area.
The following dialog box appears.
[Comment]
Comments can be added to the QC results selected by the cursor.
[None]
Select this option if you do not want to add any comments to the selected data or
you want to delete comments added previously.
[Yes:]
Select this option when you want to add comments. Up to 40 characters can be
entered in the field.
3 Click [OK].
11.8.5 Displaying the vial line
The vial line can be displayed on the L-J chart to indicate when the vial was changed.
1 Move the cursor to the first results of the date/time after changing to a new
vial.
2 Click
(Vial) in the command buttons area.
The vial line appears to the left of the cursor data.
To clear the vial line, repeat the procedure.