Chapter 11

Panel QC

3 Enter the [QC File Name] and select a panel from the [Panel] drop-down menu.
A list of subset panels is displayed as shown below.

If the panel has more than one corresponding subset panel, all of the subset panels are listed with

on the left.

Note:
If there are any unmeasured subset panels under the panel, the panel name does not appear in the dropdown menu and you cannot select it.

4 Enter the lot number for the QC file in the [Lot No.] field.
Up to 20 alphanumeric characters can be entered.

5 Confirm that the subset panels that need to be monitored by panel QC are
listed with

on the left.

If there is a subset panel that does not need to be included in the panel QC, clear the checkbox.
Only the subset panels with
are measured and monitored in daily panel QC.

6 Double-click the row to activate the subset panel you want to configure.
is shown in the [Active] column.
If there are multiple subset panels, each subset panel needs to be configured respectively.

7 Enter the expiration date for the QC material in the [Expiration Date] field.
Click
and select the expiration date of the QC material. You can also enter the date manually. You can delete
the date by clicking . If the entered expiration date has already passed, [Exp. Date] in the subset panel list is
shown in red.
You can also enter comments in the [Comment] field (optional).

136/312
XF-1600 Instructions for Use