Chapter 11

11.3

Panel QC

Panel QC files

A [QC File] is created for each panel to define the QC items and the baselines, then monitor the daily panel QC
results. It can contain the following information:
• Panel/Subset panel information
• Channels/Parameters and their baselines of Gain/PMT voltage, compensation, and/or statistics values
• Lot number and expiration date of QC material
• Limit information
• Daily panel QC results displayed on radar charts, L-J charts, and plots

Note:
Up to 50 panel QC files can be saved. One panel QC file can contain up to 200 items, and a maximum of
200 data points can be recorded for each item. After 200 data points are reached, each new data point
automatically deletes the oldest data point.

11.4

Panel QC screen

Click [QC] in the screen transition area, then select [Panel QC & Setup].
The worklist and measurement condition setting area on the main screen change to the panel QC screen below.

Panel QC file list area

Panel QC charts area

131/312
XF-1600 Instructions for Use