Chapter 10
Registering panel/subset panel and workitems
2 Enter the [Primary ID].
Alternatively, you can scan the barcode on the sample tube with the handheld barcode reader, if the
barcode label meets the requirements.
For details on acceptable barcode specifications, see the following:
(➤P.42 "7.2.1 Acceptable barcodes for sample tubes")
3 Enter the [Patient Information] and [Measurement Information].
4 Select the panel from the [Panel Name] drop-down menu.
The [Subset Panel Name] and [Subset Panel
Comment] are displayed in the [Subset Panel
Information] area.
5 Check the displayed information and enter or correct the [Secondary ID]
as necessary.
• If [Automatic] is selected for [Workitem]-[Secondary ID input] in the IPU settings:
The [Secondary ID] field displays the ID automatically assigned. If you want to change it, reenter the ID.
• If [Manual] is selected for [Workitem]-[Secondary ID input] in the IPU settings:
Enter the secondary ID in the [Secondary ID] field.
For details on the IPU setting, see the following:
(➤P.245 "Chapter 15: 15.4.10 Workitem")
Note:
• Alternatively, you can scan the barcode on the sample tube with the handheld barcode reader, if the
barcode label meets the requirements.
For details on acceptable barcode specifications, see the following:
(➤P.42 "7.2.1 Acceptable barcodes for sample tubes")
• You can also enter a [Workitem Comment], as necessary.
6 Make sure that the subset panel name for the primary ID is selected with
in the [Target] column.
If there is an unnecessary subset panel, deselect it. By default, all subset panels are selected.
Note:
Deselecting a [Target] column does not cause the secondary ID or workitem comment to be deleted.
102/312
XF-1600 Instructions for Use