Chapter 17

Maintenance

15 Plug in the power cable.
For more information about the power cable plug positions, see Chapter 3.
(➤P.27 "Chapter 3: 3.1 Analyzer")

16 Turn on the main power switch of the analyzer.

Main power switch

17 Turn on the instrument power.
For the procedure to start up the instrument, see Chapter 8. (➤P.48 "Chapter 8: 8.2 Startup")

18 Reset the air pump cycle counter.
Display the [Counter] ([Unit]) dialog box, and then click [Air Pump Reset]. For details, see the following:
(➤P.279 "17.11 Checking the cycle counter")

289/312
XF-1600 Instructions for Use