Chapter 17

Maintenance

5 Tighten the liquid waste cap by turning it clockwise.
6 Attach the cap from the empty container to the full waste container.
This step is not necessary if the full waste container has been emptied and reused.

7 Click [Accept] in the [Help] dialog box.
The [Help] dialog box closes and the status indicator LED turns green.

Risk of infection
Immediately dispose of the waste fluid according to local laws and regulations.

Caution
Install the waste container below the bottom of the instrument.

17.4

Executing auto rinse

Rinsing of the analyzer and a background check after rinsing can be executed as a single operation.

1 Display the maintenance menu.
(➤P.264 "17.2 Displaying the maintenance menu")

2 Click [Auto Rinse].
The maintenance name appears in the status area. If you
click it, the dialog box on the right appears. Progress is
shown in the progress bar.

[Auto Rinse] dialog box

Wait until the process finishes. The dialog box will automatically close and a background check will start.
For details on background checks, see the following:
(➤P.50 "Chapter 8: 8.2.3 Executing an analyzer self-test")

267/312
XF-1600 Instructions for Use