Chapter 16

Viewing logs

⏺ [Maintenance Log] screen
Displays a log of maintenance tasks executed and related information.
A maximum of 5,000 entries are stored and displayed in the maintenance log.
[Date]

Displays the date and time the log was registered.

[Logon Name]

Displays the name of the user who was logged on when the log was registered.

[Maintenance]

Displays the name of the executed maintenance task.

[Maintenance Property] Displays the property of the executed maintenance task.
[Comments]

Displays entered comments.
If you want to add a comment, select the entry and double-click or press the Enter
key to display a dialog box that allows you to enter the comment.
You can enter up to 50 characters. Once a comment is entered, it cannot be edited
or deleted. If a comment is entered when another comment already exists, the new
comment is added after the existing comment.

A maintenance log is registered when one of the maintenance tasks below is performed.
The [Maintenance] and [Maintenance Property] items that appear are below.
[Maintenance]

[Maintenance Property]

[Auto Rinse]

[As needed]

[Cleaning]

[As needed]

[Shutdown]

[Daily]

[Drain Waste Fluid Chamber]

[As needed]

[Drain Wash Chamber]

[As needed]

[Remove Flowcell Air Bubbles]

[As needed]

[Rinse Flowcell]

[As needed]

[Reagent Replenishment]

[As needed]

261/312
XF-1600 Instructions for Use