Chapter 15
15.4.15
User settings
PS-10
Click [Connect] in the [IPU Settings] tree of the [IPU Settings] dialog box, and click [PS-10].
[PS-10 Connection]
Select to enable connection to the PS-10.
This must be selected to configure connection settings.
[TCP/IP Connection]
Configure the settings to connect to the PS-10 by TCP/IP.
[PS-10 IP Address]
Set the IP address of the PS-10.
[Port No.]
Set the port number of the PS-10.
A value from 0 to 65535 can be entered.
[Communication
Protocol]
Select the TCP/IP communication format from the drop-down menu.
[Timeout]
Set the timeout for communications with the PS-10.
The minimum and maximum settings are 5 sec. and 120 sec., respectively.
The default setting is 30 sec.
Information
Before transferring data from the PS-10 to the XF-1600, make sure that the "Panel" and "Test" names for
the secondary tubes from the PS-10 are registered under the [Panel Master] tab of the XF-1600 IPU.
The "Panel" / "Test" names from the PS-10 must match the [Panel Name] / [Subset Panel Name]
respectively on the XF-1600 IPU.
If either name does not match between the XF-1600 and the PS-10, the data from the PS-10 will not be
successfully imported into the XF-1600.
249/312
XF-1600 Instructions for Use