Chapter 15

15.4.15

User settings

PS-10

Click [Connect] in the [IPU Settings] tree of the [IPU Settings] dialog box, and click [PS-10].

[PS-10 Connection]

Select to enable connection to the PS-10.
This must be selected to configure connection settings.

[TCP/IP Connection]

Configure the settings to connect to the PS-10 by TCP/IP.

[PS-10 IP Address]

Set the IP address of the PS-10.

[Port No.]

Set the port number of the PS-10.
A value from 0 to 65535 can be entered.

[Communication
Protocol]

Select the TCP/IP communication format from the drop-down menu.

[Timeout]

Set the timeout for communications with the PS-10.
The minimum and maximum settings are 5 sec. and 120 sec., respectively.
The default setting is 30 sec.

Information
Before transferring data from the PS-10 to the XF-1600, make sure that the "Panel" and "Test" names for
the secondary tubes from the PS-10 are registered under the [Panel Master] tab of the XF-1600 IPU.
The "Panel" / "Test" names from the PS-10 must match the [Panel Name] / [Subset Panel Name]
respectively on the XF-1600 IPU.
If either name does not match between the XF-1600 and the PS-10, the data from the PS-10 will not be
successfully imported into the XF-1600.

249/312
XF-1600 Instructions for Use