Chapter 2

2.3

Safety information

Installation
Warning
• Your authorized local Sysmex representative will unpack, install, and test the initial operation of the
instrument.
• This instrument must not be connected to a power outlet rated at anything other than that specified on
the rating plate. Please note that the instrument must be grounded.
Failure to do so may result in fire or electrical shock.
• Switch off the power supply before connecting any peripheral devices.
Failure to do so may result in electrical shock or damage to the instrument. In addition, an abnormal
stoppage may occur if a device is connected while the instrument is running.

Caution
• Install in a location where water will not splash or spray onto the instrument.
• Install in a location where the instrument will be protected from high temperature, humidity, dust and
direct sunlight.
• Do not install in a location subject to vibration.
• Do not subject the instrument to intense shock or vibration.
• Install the instrument in a well-ventilated area.
• Avoid installing the instrument near equipment that emits electrical interference, such as a radio or
centrifuge.
• Do not install the instrument near an area where chemicals are stored or gases are emitted.
• Do not use the instrument in a location where electroconductive gases, flammable gases, or anesthetics
that contain oxygen, hydrogen, or other flammable gases are present.
• Install the instrument indoors.
The instrument is intended for indoor use only.
• Place the reagent container at a level no more than 1 meter above or below the bottom of the analyzer.
Do not place reagents on top of the instrument.

16/312
XF-1600 Instructions for Use