Chapter 10
10.10
Registering panel/subset panel and workitems
Validating workitems
When a workitem is validated, the analysis data is approved and output to the host computer or your printer.
1 Click the [Worklist] tab in the worklist area.
2 Select the workitem that you want to validate from the worklist.
Caution
Confirm the measurement results before validating workitems.
Note:
If a different workitem is activated in the worklist at the same time, the workitem to be validated is the
selected one, not the activated one.
3 Click
(Validate) in the worklist commands.
The
icon appears in the [Validate] column.
If the selected workitem is already validated, validation is canceled and the
icon disappears.
Note:
• If hardware or measurement errors occur during the measurement of validated workitems, the error
information will be included in the data (saved in FCS format or exported to the host computer) or in the
report. For details on these errors, see the following:
(➤P.297 "Chapter 18: 18.3 Causes of errors and remedial actions")
• After validating, you cannot edit the workitem or create a report. If you need to change any information,
click
(Validate) to cancel the validation.
10.11
Reloading workitems
You can revert workitem data to its state immediately after measurement by reloading workitems. You can also revert
the data to its state at the time of the last validation operation by canceling the workitem validate operation.
1 Click the [Worklist] tab in the worklist area.
The worklist appears.
2 Select the workitem that you want to reload, and click
(Active).
115/312
XF-1600 Instructions for Use